Galvanic ProTech 1000 Z2 Tape-Based H2S Analyzer
Measured parameters: H₂S (Hydrogen Sulfide), Total Sulfur
Measurement method: Tape-Based Measurement
ProTech 1000 Z2 is a Zone 2 analyzer that measures H2S and total sulfur in natural gas by a tape-based method; sold and serviced in Türkiye by Tetra.
Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.
Technical information
- Measured parameter
- H2S (ASTM D4084); total sulfur with TS option
- Measurement range
- 0-100 ppb, 0-300 ppm, seyreltmeyle 0-100%
- Lower detection limit
- 5 ppb
- Communication
- Modbus TCP/IP and RTU, 4 x 4-20 mA
- Certificates
- CSA/UL Class I Zone 2 and Div 2, IECEx Zone 2, IP68 / Type 4X
ProTech 1000 Z2 is a tape-based analyzer that measures hydrogen sulfide (H2S) according to the ASTM D4084 method and true total sulfur with the TS option. It is used for monitoring natural gas quality at compressor stations, metering buildings and existing plant retrofits. Multi-stream measurement is possible on up to four lines for H2S and three for total sulfur.
ProTech 1000 Z2 features
- Low range 0-100 ppb, standard 0-300 ppm, 0-100% with dilution
- Accuracy and repeatability ±2% FS
- Measurement time for alarm under 20 seconds
- Modbus TCP/IP or RTU, 4 pcs 4-20 mA outputs, 4 relays
- Web-based configuration software
Spare parts and consumables
Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.
Ask about spare parts →Products from Galvanic Applied Sciences