NEO Monitors LaserGas II MP Cabinet Laser Gas Analyzer
Measurement method: TDLAS Laser Absorption
LaserGas II MP, with its multi-pass optical cell and cabinet, measures O2, CO, CO2, CH4 and NH3 at ppm level. Sold and serviced in Türkiye by Tetra.
Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.
Technical information
- Measurement principle
- Multi-pass TDLAS
- Optical path length
- 2.7 or 11.4 m
- Response time
- < 20 s (dependent on sample flow)
- Repeatability
- 1% of range
- Operating temperature
- 0 to +55 °C
- Outputs
- 4–20 mA (1–3), 3 relays; optional TCP/IP, Modbus
- Supply
- 100–240 VAC or 18–36 VDC
LaserGas II MP (Multi Pass) is a multi-pass TDLAS analyzer housed in a cabinet, used when low concentrations must be measured. A mirror system lets laser light travel an optical path of 2.7 or 11.4 m inside the sample cell, so ppm-level components can be measured in a short cell volume. The sample is brought to the cell through an extractive line.
MP configurations include O2, CO, CO2, CH4 (at ppm and % level, including CH4 in hydrogen) and H2 cabinets. ATEX Zone 2 Ex-n cabinet options can also be offered with a heated probe and heated line as a complete flue gas analysis system. Extractive furnace oxygen analysis systems are among the typical applications of this family.
LaserGas II MP features
- Multi-pass optics: 2.7 or 11.4 m optical path
- Response time shorter than 20 seconds (dependent on sample flow)
- Repeatability: 1% of measurement range
- 1–3 pcs 4–20 mA outputs, 3 relays (high gas, maintenance/warning, fault)
- Optional TCP/IP, Modbus and fiber optic communication
- 100–240 VAC or 18–36 VDC supply
- ATEX/IECEx-approved Ex-n cabinet options
Spare parts and consumables
Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.
Ask about spare parts →Products from Neo Monitors
Let's choose the right product together
Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.