Online analyzer measuring the cloud point of fuels such as diesel on the process line, used in refining and blending. Sold and serviced in Türkiye by Tetra.

Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.

Technical information

Measurement principle
Optical (light scattering) crystal detection in a Peltier-cooled cell
Compatible methods
ASTM D2500, D5771, D5772, D5773
Typical sample
Diesel fuel
Versions
Standard and Cryo (very low temperature)

The cloud point is the temperature at which wax crystals begin to form in fuel. The Cp analyzer encloses a small amount of sample in a low-mass cell and cools it at a controlled rate with a Peltier cooler. A non-contact light source and detector monitor the light scattered by crystals; when sufficient scattering is seen, the value is recorded, the cell is flushed and the cycle repeats.

Cloud Point (Cp) features

  • Results compliant with ASTM D2500 and D5771/D5772/D5773 methods; repeatability is targeted with a detection algorithm and variable-rate cooling.
  • Patented vacuum insulation improves cooling and in most cases requires no external chiller.
  • If the sample arrives very cold, the Peltier runs in reverse to heat it.
  • There are standard and Cryo versions for very low temperatures.
  • A typical application is diesel blending and adjusting additive dosing at refineries.

Choosing Cloud Point (Cp)

Because the cell is vacuum-insulated, faults from condensation and icing are reduced. The measured value is used to control additive dosing by comparison with the quality limits for winter diesel; line design matters for keeping the sample representative.

Tetra plans the sampling point and line design according to the refinery stream, commissions the analyzer and supports comparative verification against laboratory results. Service and spare parts for the cell flush cycle and Peltier module are managed from Türkiye.

Spare parts and consumables

Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.

Ask about spare parts →
CONTACT

Let's choose the right product together

Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.