Keco Model 204 Hydrocarbon / VOC in Water Analyzer

← Keco
Keco Model 204

Measured parameters: THC / VOC (Hydrocarbons)

Keco Model 204 continuously monitors hydrocarbons and VOCs in cooling tower water and wastewater; sold and serviced in Türkiye by Tetra.

Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.

Technical information

Measurement principle
Membrane stripper + tin oxide sensor or GC
Range
From low ppbw to saturation
Output
4-20 mA, RS-485 Modbus
Certificate
ATEX, IECEx (optional)

In Model 204 PermaStream, liquid sample flows continuously into a heated transfer stripper; based on Henry's law, volatile hydrocarbons pass into carrier air. The air carries them to a tin oxide sensor or gas chromatograph. Aliphatic and aromatic compounds, including BTEX, are measured; results are given in ppb, ppm or up to saturation level.

  • Early leak warning in cooling tower and heat exchanger waters.
  • Monitoring produced water, wastewater and oil-water separator outlets.
  • Stormwater, groundwater recovery points and condensate water.
  • Continuous online analysis requiring no routine field calibration.
  • Liquid verification with optional PermTube and automatic verification system.
  • 4-20 mA, RS-485 Modbus, high/low alarm and sampling system diagnostic alarm.
  • Hydrocarbon emission control unit and H2S scrubber options.

Because measurement is made by directly stripping free or dissolved hydrocarbon in water from the sample, it is important for the liquid sample to stay representative along the line. The sample point, line diameter and particle quantity should therefore be reviewed before installation.

Tetra assesses particle size, flow and temperature in the sample line and plans the pretreatment for Model 204. Periodic service visits are made for stripper cleaning, sensor checks and verification.

Spare parts and consumables

Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.

Ask about spare parts →
CONTACT

Let's choose the right product together

Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.