M&C ECM Series Compressor Gas Sample Cooler
M&C ECM series compressor gas coolers dry sample gas for CEMS and process analyzers. Sold and serviced in Türkiye by Tetra.
Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.
Technical information
- Series
- ECM-1, ECM-2, ECM-Ex2 (parts compatible with ECP / ECC exchangers)
- Exchanger material
- Duran glass, stainless steel, PVDF (option)
- Supply voltage
- 240 V or 120 V, 50-60 Hz
- Connections
- Sample gas GL 18-6 mm, condensate GL 25-12 mm
The M&C ECM series are compressor gas coolers that lower the dew point of sample gas on flue gas measurement (CEMS) and process gas analysis lines. Separating moisture as condensate before it reaches the analyzer protects the analyzer cell and allows measurement on a dry gas basis. They are used with a heated probe and heated line in applications such as thermal power plants, cement and chemical plants, waste incineration and refineries.
The series covers ECM-1 (one exchanger), ECM-2 (two exchangers) and ECM-Ex2 models for Zone 2 hazardous areas. Jet-stream exchangers shared with the family's ECP and ECC type coolers (with material options such as Duran glass, stainless steel and PVDF) and components such as the fan, temperature controller and PT100 sensor are offered as separate items.
ECM features
- On the ECM-1 data sheet: 1 jet-stream exchanger, maximum gas flow 250 Nl/h, outlet dew point stability better than ±0.1 °C under constant conditions
- ECM-1 total cooling capacity 144 kJ/h at +10 to 50 °C ambient temperature; ambient temperature up to +50 °C
- ECM-2: ready for two jet-stream exchangers (150 Nl/h each), outlet dew point factory-set to +5 °C, high/low temperature alarm and digital thermometer
- Configurable status alarm contact; compact wall-mounted housing
- Optional peristaltic drain pump (SR25.2) and different gas/condensate connections depending on exchanger type (GL 18-6 mm, GL 25-12 mm)
- ECM-Ex2 (single or dual channel) applications for Zone 2
For your single- or multi-channel application, simply share the sample flow, ambient condition and hazardous area requirement; Tetra will select the suitable exchanger and pump.
Spare parts and consumables
Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.
Ask about spare parts →Products from M&C TechGroup
Let's choose the right product together
Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.