SENTRY DCCP Degassed Cation Conductivity Panel
Measured parameters: Cation Conductivity
Measurement method: Conductometric Conductivity
The SENTRY DCCP degas panel makes degassed cation conductivity measurable in the steam-water cycle. Sold, installed and serviced in Türkiye by Tetra.
Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.
Technical information
- Measured parameter
- Degassed cation conductivity
- Supply voltage
- 110-120 VAC or 220-240 VAC, single phase, 50/60 Hz
- Resin column
- RC-100
- Backplate
- Flat (DCCP-124101)
SENTRY DCCP (Degassed Cation Conductivity Panel) is a panel system prepared to measure degassed cation conductivity in boiler water and steam samples at power plants. The sample first passes through a cation resin column, and dissolved gases (particularly CO₂) are then removed, so the conductivity measurement reflects salt and acid contamination more accurately. It is used in steam turbine and boiler feedwater chemistry monitoring.
The panel may include an RC-100 resin column, flow indicator and valve, as well as a conductivity analyzer kit (Thermo Scientific 2100 UPW conductivity analyzer kit). The W7950 degas sparger with sample-valved flow meter is also offered separately in this family.
DCCP features
- 110-120 VAC or 220-240 VAC single-phase supply (degas panel)
- DCCP-124101: flat backplate version
- RC-100 resin column can be integrated
- Sample flow control with flow indicator and valve
- As a complete measurement system with the Thermo Scientific conductivity analyzer kit
If you state the number of sample points at your plant and your conductivity analyzer preference, Tetra will configure the panel.
Spare parts and consumables
Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.
Ask about spare parts →Products from SENTRY Equipment
Let's choose the right product together
Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.