SENTRY TSV Thermal Shut-Off Valve

← SENTRY Equipment
SENTRY TSV

SENTRY TSV is a mechanical valve that cuts flow when sample temperature exceeds the set value, protecting analyzer sensors. Sold and serviced in Türkiye by Tetra.

Installation, calibration, spare parts and service in Türkiye come from Tetra; cabinet and panel integration can be done at our İzmir facility.

Technical information

Operating principle
Mechanical thermal shut-off, manual reset
Set points
43, 49, 55, 60, 67, 71, 95 °C

SENTRY TSV (Thermal Shut-Off Valve) is used to protect analyzer sensors on the sample line from high temperature. If the sample temperature on the panel exceeds the set value, this mechanically operated valve automatically cuts off sample flow. This protects the cooler, flow indicator, BPRV and sensors from damage, prevents steam discharging to atmosphere and ensures operator safety.

Typical causes of a temperature rise are loss of cooling water, high sample flow rate, high cooling water temperature, a dirty or clogged sample cooler and operator error. The valve requires a manual reset and has a visual trip indicator.

TSV features

  • Mechanical, unpowered automatic shut-off valve
  • Listed set points: 43, 49, 55, 60, 67, 71 and 95 °C (depending on model code)
  • Requires manual reset; visual trip indicator
  • Helps protect the secondary cooling system (TCU) from overload
  • W-270 HTSO high-temperature shut-off valve option

If you state the shut-off temperature you want, Tetra will supply a TSV with a suitable set point.

Spare parts and consumables

Write to us for spare parts and consumables; the Tetra service team supplies and replaces the right original parts for your instrument.

Ask about spare parts →
CONTACT

Let's choose the right product together

Tell us about your application; we will recommend the right model and configuration for the measuring point, process conditions and regulations.